Nominal voltage
or
Frequency
Heating function
Heating output
Temperature range (medium)
Temperature range (medium)
Setting accuracy
Controller deviation (3 l water/90 °C)
Absolute deviation / mean (3 l water/90 °C)
Controller deviation (3 l Marlotherm oil/150 °C)
Absolute deviation / mean (3 l Marlotherm oil/150 °C)
Stirring function
Motor
Motor output
Speed range
Minimum filling height
Overvoltage class
Permissible operating time
Protection class according to DIN EN 60529
Permissible ambient temperature
Permissible relative humidity
Protection class
Total volume - bath container
Material - bath container
Dimensions (Ø x H)
Weight
Interfaces
Subject to technical changes!
Warranty
The warranty does not cover wearing parts, nor to defects that are
the result of improper use, insufficient care and maintenance or
failure to follow the instructions in this operating manual.
Technical data
VAC
VAC
Hz
W
°C
K
K
K
K
K
W
rpm
cm
%
°C
%
l
mm
kg
HBR 4 control
230 ± 10%
115 ± 10%
50/60
1000
Room temperature...200
Set and actual temperature LCD
1
± 0.4
± 1
± 0.8
± 2
Ballbearing brush-free, EC motor
5
150...800 adjustable in steps of 50 rpm
3
II
100
IP 20
+5...40
80
DIN 12 877 class 2
5
1.4301
340 x 250
4.4
PT 1000.60/61
RS 232
HBR 4 digital
1
± 0.4
± 1
± 0.8
± 2
5
3
5
340 x 250
4.4
-
25