Stanford Research Systems DS360 Operating Manual And Programming Reference page 71

Ultra low distortion function generator
Table of Contents

Advertisement

DIGF (?) i
STPE (?) i
FSTP (?) x
DTAA (?) x
DTBA (?) x
DTAF (?) x
DTBF(?) x
The DIGF command sets (queries) the digital sampling frequency for the digital
output, based on the table below.
i
0
1
2
The STPE command enables (i=1) or disables (i=0) the usage of the frequency
step, which is set by the FSTP command. The STPE? query returns the current
step enable mode.
The FSTP command sets the frequency step to x Hertz. The FSTP? query returns
the current step frequency. The step is active only for the main frequency
(FREQ) and STPE (step enable) = 1. The value x may range between 10mHz and
200kHz.
When the frequency step is enabled, it becomes the knob increment value.
The DTAA command sets the Tone 1 digital amplitude to x percent (%). The
maximum value for the sum of Tone 1 and Tone 2 amplitude is limited as
described in chapter 5. The DTAA? query will return the amplitude in percent
(%).
The DTBA command sets the Tone 2 digital amplitude to x percent (%). The
maximum value for the sum of Tone 1 and Tone 2 amplitude is limited as
described in chapter 5. The DTBA? query will return the amplitude in percent
(%).
The DTAF command sets the Tone1 digital frequency to x Hertz. The DTAF?
query returns the current Tone1 frequency. The frequency is set and returned
with 10mHz resolution.
The DTBF command sets the Tone 2 digital frequency to x Hertz. The DTBF?
query returns the current Tone 2 frequency. The frequency is set and returned
with 10mHz resolution.
Remote Programming 4-13
Frequency
48.0 kHz
44.1 kHz
32.0 kHz
DS360 Ultra Low Distortion Function Generator