Guidance And Manufacturer's Declaration - Electromagnetic Emissions - Wincomm WMP-19B User Manual

Medical panel pc
Table of Contents

Advertisement

Guidance and manufacturer's declaration – electromagnetic
emissions
The model WMP-19B/22B is intended for use in the electromagnetic
environment specified below. The customer or the user of the model
WMP-19B/22B should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Recommended separation distances between
portable and mobile RF communications equipment and the model
The model WMP-19B/22B is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The
customer or the user of the model WMP-19B/22B can help prevent
electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment
(transmitters) and the model WMP-19B/22B as recommended below,
according to the maximum output power of the communications
equipment.
Rated maximum
output power of
transmitter
W
0,01
Compliance
WMP-19B/22B
Separation distance according to
150 kHz to
80 MHz
d = 1,2
0,12
WMP-19B/22B User's manual
Electromagnetic environment –
guidance
The model WMP-19B/22B uses RF
energy only for its internal function.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
The model WMP-19B/22B is suitable
for use in all establishments,
including domestic establishments
and those directly connected to the
public low-voltage power supply
network that supplies buildings used
for domestic purposes.
frequency of transmitter
m
80 MHz to
800 MHz
d = 1,2
0,12
800 MHz to
2,5 GHz
d = 2,3
0,23