2.0 Transmitter installation

2.1 Mount the transmitter

Liquid applications
1. Place taps to the side of the
line.
2. Mount beside or below the
taps.
3. Mount the transmitter so that
the drain/vent valves are
oriented upward.
Gas applications
1. Place taps in the top or side of
the line.
2. Mount beside or above the
taps.
Steam applications
1. Place taps to the side of the
line.
2. Mount beside or below the
taps.
3. Fill impulse lines with water.