Agilent Technologies 89410A Operator's Manual page 280

Hide thumbs Also See for 89410A:
Table of Contents

Advertisement

Digital Demodulation Concepts (Opt. AYA)
Another way of viewing the digital demodulation process is illustrated in the
following table:
Measurement Path for Digital Demodulation
Parameters
Time Sequence
Involved
Signal setup
Range
Trigger
Frequency/span
Input
Time Corrections
Window
Memory
Pre-demodulation
Pulse search
Search length
Result length
Demodulation
Setup parameters
Format
Symbol rate
Filter
Result length
Sync search
Points/symbol
Clock adjust
Post-demodulation
Measurement data
Display format
Scale
Marker
Normalize
22 - 6
Effect
Result
Sets up basic signal
Raw data
parameters required
for digital
demodulation
Determines what
Input to
signal enters the
demodulator
demodulator by
trimming the raw
data
Determines the
Demodulated,
type of
symbol-locked data
demodulation
in the default
performed on the
display format
input time data
Formats the data
Formatted,
for display
demodulated data
Notes
Any changes made
in the signal setup
parameters later in
the time sequence
require that a new
measurement be
performed
This signal is the
maximum possible
length, limited only
by memory
constraints.
The result length
derived previously
may be shortened if
desired
Changes may be
made in the display
parameters without
acquiring a new
measurement