To View Data For One Or More Slots - Agilent Technologies 89410A Operator's Manual

Hide thumbs Also See for 89410A:
Table of Contents

Advertisement

Using Wideband CDMA (Options B73, B79, and 080)

To view data for one or more slots

This task builds on the previous task to show you how to use time gating.
Time gating lets you view measurement data for selected slots.
1.
Perform the previous task.
2.
Display the IQ magnitude error:
Press [
].
B
Press [
Measurement Data
3.
Autoscale the results:
Press [
].
Auto Scale
4.
Display the IQ magnitude error for the first slot:
Press [
], [
Time
gate length],
Press
0
[gate delay],
, [slot].
Changing [
or
gate length]
gating.
5.
Display the IQ magnitude error for the second slot:
Press
[gate delay]
Press the up arrow key (to increment gate delay to 1 slot).
6.
Display the IQ magnitude error for the second slot in the previous code channel:
Press
[ code channel] .
Press the down arrow key (to decrement code channel to code 9).
This task introduces time gating, which is an advanced feature that lets you
select one or more slots for analysis. With time gating off, the
measurement includes all slots (as set by [
With time gating on, the measurement includes the slots selected by
and
Gate length selects the number of slots; gate delay
length]
[gate delay].
determines which slot in the measurement to use as the starting slot. For
example, a gate delay of zero selects the first slot in the measurement as
the first slot in the gated results; a gate delay of one selects the second
slot; a gate delay of two selects the third slot, and so forth.
13 - 10
], [
].
time domain], [IQ error mag
1
, [slot].
automatically selects [
[gate delay]
Time
] to enable time
gate on
], [
]).
main length
[gate