Agilent Technologies 89410A Operator's Manual page 183

Hide thumbs Also See for 89410A:
Table of Contents

Advertisement

Press [
] 22 [
maximum freq
3.
To enable control of the external spectrum analyzer via GPIB:
rcvr control on]
Press [
Press [
], [
Local/Setup
(or use whatever address the attached RF/microwave spectrum analyzer is
set to).
4.
Set up the other measurement parameters.
Press [
Instrument Mode
Press [
], [
System Utility
Press [
], [
center
Frequency
Press [
], 3 [
span
MHz
], −14 [
Press [
Range
Press [
], [
ResBW/Window
Press [
], [
main window
Press [
], [
Return
num freq pts
5.
Set up the source and turn it on:
Press [
], [
source type
Source
], −22, [
Press [
], [
level
Source
Press [
].
source on
6.
Set up the trigger:
Press [
], [
trigger type
Trigger
7.
Set up averaging.
Press [
], [
average type
Average
Press [
], [
Return
num averages
Press [
].
average on
8.
Set up a time display:
Press [
Measurement Data
Take a measurement: press [
Wait for the average count to reach 100.
Store this data in D3: press [
Turn averaging off: press [
(continued on next page...)
] (appropriate for the HP/Agilent 8566A/B).
GHz
], [
system controller
peripheral addresses
], [
].
vector
].
time domain cal on
], 5.6 [
].
MHz
].
].
dBm
].
rbw mode arb
].
uniform
], 801, [
].
Enter
], [
].
periodic chirp
].
dBm
], [
].
internal source
], [
].
time
] , 100, [
].
enter
],
[main time
].
ch1
].
Meas Restart
], [
save trace
Save/Recall
], [
average off
Average
Using the Agilent 89411A Downconverter
], [
], 18, [
ext rcvr adrs
], [
].
into D3
].
]
Enter
15-9