The signal of 4 mA always corresponds to a measured value of 0 mg/m³.
"Full-Scale" always denotes the residual oil content at a full-scale deflection of 20 mA.
"Resolution" indicates the resolution of the interface.
Example setting 1
If the residual oil content values indicated in the METPOINT OCV
mg/m³, please enter "0.001" into the blue field and confirm with OK.
After confirmation with OK, the value for "Full-Scale" and the resolution are indicated. In this example, the following
settings are applicable:
Scaling
Scaling initial value
Scaling final value (Full-Scale)
METPOINT
®
OCV
0.1 mA
corresponds to
4.0 mA
corresponds to
20 mA
corresponds to
®
MEASUREMENT mode lie in the range of ≤ 0.1
0.001
mg/m³
0.000 mg/m³
0.160
mg/m³
Operation
63