10.7 Installation steps

The METPOINT
personnel, and exclusively in accordance with the technical data.
For the connection to the sensor unit, do not touch the probe without wearing oil- and grease-free gloves
(textile gloves are recommended). A fingermark on the probe or in the measuring zone can affect the
measuring results or lead to false measurements.
Prior to installation, allow compressed air to flow off for a short period of time to remove condensate and
particles. This will prevent the unnecessary contamination of the probe and of the sensor unit.
NOTE!
Strong variations of the ambient temperature, also during transport and storage can lead to damage to the
device or to incorrect measuring results.
Please make sure that the METPOINT
prior to installation at the place of installation so as to allow the temperature adaption to the ambient
temperature. During this period, the sensor unit must not be connected to the power supply. Careful
observance of this point is imperative in particular in the case of major temperature differences between the
transport / place of storage and the place of use.
The reference gas generation unit requires a pre-heating time of approx. two hours. Only after this pre-
heating time has expired may compressed air flow through the METPOINT
waiting time, the sensor unit can be connected to the power supply at this time.
Unpacking of the METPOINT
METPOINT
®
OCV
®
OCV measuring devices must only be installed and used by qualified and authorised
®
OCV (steps 1 – 3)
®
OCV device is unpacked and stored approximately three hours
1. Please wear clean (oil-and grease-free) gloves during the
entire installation of the METPOINT OCV
2. Please take the probe (3) of the METPOINT
packaging and make sure that you do not yet remove the
sheet package of the sampling probe.
3. Please keep the supplied documentations in a safe place.
Installation
®
OCV device. To reduce the
®
.
®
OCV from the