Beko METPOINT OCV Installation And Operation Manual page 21

Oil vapour measuring device with sampling probe
Table of Contents

Advertisement

Basic requirements regarding the installation:
The measuring gases must be free from dust and water
when supplied to the METPOINT
Therefore, gas processing adjusted to the measurement task
is required in most application cases.
Choose a clean, dry and dust-free room which is protected
against storm.
The METPOINT
hazardous area!
The measuring section must meet the following
requirements:
- centric probe seating
- made of stainless steel
- oil- and grease-free
- dimensions in accordance with the BEKO
provisions
- inflow towards the probe from the bottom to the top
(see flow direction)
The measuring section must be installed vertically.
The employment of an original BEKO measuring
section is recommended.
The installation of a bypass (see Chapter 10.5) or of
a ball valve between the sensor unit and the
measuring section is recommended.
Ambient temperature (sensor unit) +5°C to 45°C.
Ambient temperature (evaluation unit with user
interface)
+5°C to +50ºC.
The maximum pressure dew point of +10°C must
not be exceeded.
Other connection possibilities and measuring
methods are possible, depending on the respective
application. Please contact us in this regard.
METPOINT
®
OCV
®
OCV measuring device.
®
OCV must not be employed in a
Installation
Measuring
gas inlet
21