Ac Power (45 Hz To 65 Hz) Specification Summary, Pf=1 - Fluke 5500A Getting Started

Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

AC Power (45 Hz to 65 Hz) Specification Summary, PF=1

Voltage Range
5500A Calibrator
90 days
33 to 329.999 mV
330 mV to 1020 V
1 year
33 to 329.999 mV
330 mV to 1020 V
5725A Amplifier
90 days
100 to 1020 V
1 year
100 to 1020 V
Voltage Range
5500A Calibrator
90 days
33 to 329.999 mV
330 mV to 1020 V
1 year
33 to 329.999 mV
330 mV to 1020 V
5725A Amplifier
90 days
100 to 1020 V
1 year
100 to 1020 V
Voltage Range
5500A Calibrator
90 days
33 to 329.999 mV
330 mV to 1020 V
1 year
33 mV to 1020 V
330 mV to 1020 V
Note
1. To determine uncertainty with more precision, see "Calculating Power Uncertainty."
5500A Calibrator Current Range
3.3 to 8.999 mA
9 to32.999 mA
Absolute Uncertainty, tcal + 5°C, + (% of watts output)
0.30%
0.20%
0.20%
0.12%
0.40%
0.25%
0.25%
0.15%
0.20%
0.12%
0.25%
0.15%
5500A Calibrator Current Range
0.33 to 0.8999 A
0.9 to 2.1999 A
Absolute Uncertainty, tcal + 5°C, + (% of watts output)
0.25%
0.20%
0.20%
0.12%
0.35%
0.25%
0.25%
0.15%
0.20%
0.12%
0.25%
0.15%
5725A Amplifier Current Range
1.5 to 4.4999 A
Absolute Uncertainty, tcal + 5°C, + (% of watts output)
0.25%
0.15%
0.35%
0.20%
Multi-Product Calibrator
33 to 89.99 mA
90 to 329.99 mA
0.25%
0.20%
0.20%
0.12%
0.35%
0.25%
0.25%
0.15%
0.20%
0.12%
0.25%
0.15%
2.2 to 4.4999 A
4.5 to 11 A
0.25%
0.20%
0.18%
0.12%
0.35%
0.25%
0.20%
0.15%
0.18%
0.12%
0.20%
0.15%
4.5 to 11 A
0.20%
0.12%
0.25%
0.15%
Specifications
25

Advertisement

Table of Contents
loading

This manual is also suitable for:

5500e

Table of Contents