Specifications

The following paragraphs describe the details for the 5500A specifications. All
specifications are valid after allowing a warm-up period of 30 minutes, or twice the time
the 5500A has been turned off. (For example, if the 5500A has been turned off for 5
minutes, the warm-up period is 10 minutes.)
All specifications apply for the temperature and time period indicated. For temperatures
outside of tcal + 5°C (tcal is the ambient temperature when the 5500A was calibrated),
the temperature coefficient is less than 0.1 times the 90-day specifications per °C
(limited to 0°C - 50°C). These specifications also assume the 5500A Calibrator is zeroed
every seven days or when the ambient temperature changes more than 5°C. (See
"Zeroing the Calibrator" in Chapter 4 of the Operator Manual.)
Also see additional specifications later in this chapter for information on extended
specifications for ac voltage and current. The dimensional outline for the 5500A
Calibrator is shown in Figure 6.
5500A
CALIBRATOR
NORMAL
AUX
SCOPE
V,
,
A,
-SENSE,
200V PK
RTD
AUX V
MAX
STBY
HI
7
1000V
20V
RMS
RMS
MAX
MAX
TRIG
LO
OUT
4
1V PK
MAX
20V PK
MAX
1
TC
/ +
20V PK
MAX
43.2 cm (17 in)
PREV
OPR
EARTH
SCOPE
BOOST
MENU
µ
dBm
sec
8
9
SETUP
RESET
m
V
Hz
n
W
¡F
NEW
5
6
k
A
¡C
CE
REF
p
MEAS
TRIG
2
3
M
F
TC
OUT
MULT
DIV
0
SHIFT
ENTER
x
÷
47.0 cm (18.5 in)
Figure 6. 5500A Calibrator Dimensional Outline
17.8 cm
(7 in)
EDIT
FIELD
POWER
I
O
Multi-Product Calibrator
Specifications
6.4 cm
(2.5 in)
For Cable
Access
F1-04.eps
11

Advertisement

Table of Contents
loading

This manual is also suitable for:

5500e

Table of Contents