NPI ELC-03XS Operating Instructions And System Description page 10

Universal amplifier
Table of Contents

Advertisement

_________________________________________________________________________________________________________________
(3) REMOTE connector
BNC connector to attach a remote switch in active-low configuration for activating
the PENETRATION / BUZZ unit.
(4) MANUAL push button
Push button for activating the PENETRATION / BUZZ unit manually. The
DURATION is set by 48 and the mode by 2.
Note: The duration is dependent on the setting of the DURATION potentiometer, but
independent from how long the button is pushed or a REMOTE switch is pressed.
potential (mV)
12000
8000
4000
0
-4000
-8000
start
-12000
Figure 2: BUZZ function of the ELC-03XS (CC Mode)
(5) FREQUENCY potentiometer
Potentiometer to set the FREQUENCY of pulses within a train in PULSE mode
(see also 2). Range: 100 Hz to 2 kHz.
(47) AMPLITUDE potentiometer
Potentiometer for setting the AMPLITUDE of pulses in PULSE mode (see also 2).
(48) DURATION potentiometer
Control to set the duration of the BUZZ (potentiometer, clockwise, range: ~1 ms to
~100 ms). It is effective in all modes; REMOTE controlled and when pushing the
MANUAL button.
___________________________________________________________________________
version 2.2
ELC-03XS User Manual
stop
time
page 10