11. Measurement practice
For
best
measurement
accuracy
keep
the
compression/tension forces in line with the force gauge.
Alleviate bending loads and torque loads applied to the
load cell as these can adversely affect measurement
performance.
Always keep the gauge below the capacity limit shown on
the front of the gauge. If gauge is used above this capacity
in either tension or compression, even for a short time,
permanent load cell damage can result. Overload damage
is not covered by warranty.
Accuracy: ± 0.2 % of rated capacity
Operating temperature: 60 ºF - 95 ºF (15 ºC - 35 ºC)
Temperature shift at zero load: ± 0.04 % of full-scale/ºC
Output
RS-232 and USB: 8 data bits, 1 Start bit, 1 Stop bit, no
parity
Baud rate: 38400
Peak Capture Rate : 0.100 S
ADC Sampling Rate : 1,000 Hz
FL-BA-e-1411
6