ChemStation Views
Verification (OQ/PV) View
Verification (OQ/PV) View
Start/Stop Run
Task
Selection
Menu
System
Diagram
Message
Line
The Verification (OQ/PV) view lets Agilent service engineers test whether your
analytical instruments and the ChemStation software are operating correctly
according to predefined performance criteria.
Operation Qualification (OQ) Operation Qualification is the documented
verification that the equipment-related system or subsystem performs as
intended throughout representative or anticipated operating ranges.
14
Verification Toolbar
Agilent 6100 Series LC/MS System Quick Start Guide