Malmet WDS Operation, Maintenance And Installation Manual page 30

Washer disinfector sanitizer
Table of Contents

Advertisement

WDS
Operation, Maintenance and Installation Manual
401
Steam Heatup
403
Steam Heatup
404
Steam Heatup
405
Steam Heatup
421
Steam Disinfect
422
Steam Disinfect
423
Steam Disinfect
801
Purge flush
802
Purge flush
803
Purge flush
804
Purge flush
805
Purge Cold Fill
806
Purge Cold Fill
ISSUE 5
Steam Water Level Timeout
Steam Water Low Level during
heatup
Steam Water Temperature High
Steam Timeout
Disinfect Water Low Level
Disinfect Water Temperature High
Disinfect Timeout
Low Water Level during wash
Door Open
Pump run without solenoid
Timeout before Level reached
Water Level Sensor Fail
Water Fill Timeout
30
Page
Water Leakage, Steam fill solenoid failed,
Steam Level Sensor Failed, Blocked inlet
hose
Water Leakage, Steam fill solenoid failed,
Steam Level Sensor Failed, Blocked inlet
hose
Faulty Temperature Sensor, Faulty Heater
Relay, Faulty contactors,
Faulty steam heater element, Faulty steam
heater relay, Heater Circuit Breaker tripped,
Overt temp switch tripped. Faulty temp
probe, Faulty contactors. Panel lockout
switches J8 open circuit, Door interlock
switch J11 J10
Water Leakage, Steam Level probe Failed,
Faulty Temperature Sensor, Faulty Heater
Relay, Faulty contactors,
Faulty steam heater element, Faulty steam
heater relay, Heater Circuit Breaker tripped,
Overt temp switch tripped. Faulty temp
probe chamber steam outlet, Faulty
contactors. Blocked condenser coil.
Blocked mesh on steam outlet, Panel
lockout switches J8 open circuit, Door
interlock switch J11 J10
Low pump water flow, blocked spray
nozzles, Faulty solenoid, Panel lockout
switches J8 open circuit. Faulty level probe.
Incorrect spray nozzles fitted
Door opened when attempting pure rest,
faulty door interlock switch J11 J10
Check Cycle Times parameters, Faulty
solenoids
Low pump water flow, blocked spray
nozzles, Faulty solenoid, Panel lockout
switches J8 open circuit. Faulty level probe.
Door interlock switch J11 J10,
Dirty Conductivity Sensor Probe, faulty
connections, faulty conductivity module,
Water Level Sensor Failed or Water supply
low flow, Water Fill solenoid failed, water
supply off.