7.8 Automatic adding-up

With this function the individual weighing values are automatically added into the
summation memory when the balance is unloaded without pressing
when an optional printer is connected.
• Menu settings:
„P1 COM"
Indicator
AUTO
Add up:
Place goods to be weighed A.
After the standstill control sounds a signal tone. The weighing value is added to
the summation memory, followed by printing.
Remove the weighed good. More weighed goods can only be added when the
display = zero.
Place goods to be weighed B.
After the standstill control sounds a signal tone. The weighing value is added to
the summation memory, followed by printing. The number of weighings, followed
by the total weight, will be shown for 2 sec.
Add more weighed goods as described before.
Please note that the weighing system must be unloaded between the individual
weighing procedures.
This process can be repeated until the capacity of the weighing system is
exhausted.
Display and delete the weighing data, as well as printout examples
see chpt. 7.7.