9 Service, maintenance, disposal

9.1 Cleaning

• Before cleaning, disconnect the appliance from the operating voltage.
• Do not apply aggressive detergents (solvents etc.).

9.2 Service, maintenance

The appliance may only be opened by trained service technicians who are authorized
by KERN.
Before opening, disconnect from power supply.

9.3 Disposal

Disposal of packaging and appliance must be carried out by operator according to
valid national or regional law of the location where the appliance is used.

9.4 Error messages

Error message
- - - - -
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
IFB-BA-e-1012
Description
Maximum load exceeded
Zeroing range exceeded
due to switching-on balance
or pressing
(normally
4% max)
Value outside the A/D
changer range
Possible causes
• Remove load from scale
or reduce preload.
• Object on the weighing
plate
• Overload when zeroing
• Improper adjustment
• Damaged weighing cell
• Damaged electronics
• Weighing plate not
installed
• Damaged weighing cell
• Damaged electronics
27