KERN IFB-BA-e-1012 Operating Instructions Manual page 14

Platform scale
Table of Contents

Advertisement

Switch on instrument and press
Press
,
,
CHK" will appear.
Press
repeatedly until "P2 CAL" appears.
Press
and the first menu item "COUNT" will appear
At verified scales, press first the adjustment switch.
Press
repeatedly until "CAL" appears.
Press
.
Wait for stability display, then press
are no objects on the weighing plate.
The currently set adjustment weight will be displayed.
Change by pressing the navigation keys (See chpt 2.1.1);
select desired setting; currently enabled digit will be flashing.
Acknowledge by
Carefully place adjusting weight in the centre of the weighing
plate. Wait for stability display, then press
After successful adjustment the appliance automatically
returns to weighing mode.
An adjusting error or incorrect adjusting weight will be indicated by the error
message, repeat adjustment procedure.