DeVilbiss iGo 306DS Instruction Manual page 10

Portable oxygen concentrator system
Hide thumbs Also See for iGo 306DS:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
SYMPTOM
Cannot adjust/change the flow rate.
Yellow Low Oxygen light is illuminated.
OR
Red Service Required light is illuminated with
audible alert.
External Power and/or Battery Power lights
illuminated. Unit is operating.
Red Service Required light is illuminated.
External Power and/or Battery Power lights
illuminated. Audible alert is sounding. Unit does
not operate.
Power Fail Alert activates (External Power light
flashes and audible alert sounds).
Red Service required light is flashing. Flow
Indicator light is flashing. Audible Alert is
sounding. Unit is operating.
Unit does not operate when using with charged
battery. Power Fail alarm activates.
DC Adapter is attached. Unit runs from Battery
or does not run. The External Power Present
Light not illuminated.
Lowest Battery Status Light is Yellow. Unit
beeps once.
Battery Status Lights never indicate fully
charged.
Yellow Battery Status light is flashing.
Audible Alert sounds intermittently when
operating from DC Adapter.
Short Battery Run Time.
If any other problems occur with your iGo.
cArInG For Your deVIlbISS IGo PortAble oXYGen SYSteM
WArnInG
Before attempting any cleaning procedure, turn the unit Off and disconnect from AC or dC power.
do not use lubricants, oils, or grease.
Air Filter
The air filter should be cleaned at least once a week. To clean, follow these steps:
1. Remove the air filter.
2.
Wash in a solution of warm water and dishwashing detergent.
3. Rinse thoroughly with warm tap water and towel dry. The filter should be completely dry before reinstalling.
CAUTION-To prevent product damage, do not attempt to operate the unit without the air filter or while the filter is still damp.
NOTe- Device has an internal intake filter that must be inspected by a qualified provider/technician every 3 years.
en - 10
POSSIBLE CAUSE
1. Flow rate has been locked by your provider.
2. Unit malfunction.
1. Air filter is blocked.
2. Exhaust is blocked.
1. Air filter is blocked.
2. Exhaust is blocked.
3. Unit is overheated.
1. Battery completely discharged.
2. Lost external power without battery installed.
1. Blocked or defective cannula or oxygen tubing
1. Battery is overheating.
1. Poor connection to DC power source.
2. Vehicle power source (vehicle accessory connector)
dropped too low for the DC Adapter.
3. Fuse in vehicle is blown.
1. Battery needs to be charged.
1. Battery needs to be conditioned.
2. Defective battery.
1. Unit is equipped with built-in battery test and the
battery pack is being checked.
1. Vehicle not running.
2. Poor connection to DC power source.
3. Vehicle electrical system overloaded or defective.
1. Battery needs to be reconditioned.
2. Air filter is blocked.
3. Exhaust is blocked.
4. Intake filter silencer is blocked.
REMEDY
1. Contact your DeVilbiss provider.
2. Contact your DeVilbiss provider.
1. Check the air filter. If the filter is dirty, wash it following the cleaning
instructions.
2. Check the exhaust area; make sure there is nothing restricting the unit
exhaust.
1. Check the air filter. If the filter is dirty, wash it following the cleaning
instructions.
2. Check the exhaust area; make sure there is nothing restricting the unit
exhaust.
3a. Allow unit to cool and try again.
3b. Move unit to cooler location.
1. Recharge battery.
2. Install battery or plug into external power.
1. Detach cannula. If proper flow is restored, clean or replace if necessary.
Disconnect the oxygen tubing at the oxygen outlet. If proper flow is restored,
check oxygen tubing for obstructions or kinks. Replace if necessary.
1. Allow battery to cool.
1. Ensure the DC accessory outlet is clean and a good connection can be
made. Insert the other end of the DC Adapter into the vehicle's DC
accessory outlet.
2. If the vehicle power source drops too low for the DC Adapter, the iGo will
revert to Battery operation (if present) or will cease operation until power is
restored.
3. Check the fuse and replace if necessary.
1. Plug unit into AC power or replace discharged battery with charged
battery.
2. Plug unit into DC accessory outlet to operate unit (battery will not
recharge on DC power).
1. Fully discharge battery then recharge.
2. Contact your DeVilbiss provider.
1. If light flashes longer than 5 minutes, battery is defective. Contact your
DeVilbiss provider for a replacement.
1. Start vehicle.
2. Ensure the DC accessory power port outlet is clean and a good
connection can be made. Insert the DC Adapter into the vehicle's DC
accessory outlet.
3. Have a qualified auto technician check electrical system while iGo is
attached.
1. Perform a complete discharge and then a 100% charge of the battery.
2. Check the air filter. If the filter is dirty, wash it following the cleaning
instructions.
3. Check the exhaust area; make sure there is nothing restricting the unit
exhaust.
4. R eplace the intake silencer/filter.
1. Turn your unit Off and switch to your reserve oxygen system. Contact
your DeVilbiss provider immediately.
A-306-1