Inspect electrical components for signs of corrosion. Replace if corroded or damaged.
Inspect battery terminals for loose cable connection. Tighten if necessary.
Inspect all fasteners.
Inspect TRBKTS fasteners and hardware.
Inspect foam handgrips for damage. If damaged, have them replaced by a qualified technician.
Ensure swingarm stops are in place and not deteriorated or damaged. Replace if necessary.
Inspect/Adjust Every 18 Months
Replace motor brushes and gearbox coupling.
11.4 Batteries
DANGER
Risk of Death or Serious Injury
Failure to observe these warnings can cause an electrical short resulting in death, serious injury, or damage to the electrical
system.
The POSITIVE (+) RED battery cable MUST connect to the POSITIVE (+) battery terminal(s)/post(s).
The NEGATIVE (-) BLACK battery cable MUST connect to the NEGATIVE (-) battery terminal(s)/post(s).
NEVER allow any of your tools and/or battery cable(s) to contact BOTH battery post(s) at the same time. An electrical
short may occur and serious injury or damage may occur.
Install protective caps on positive and negative battery terminals.
Replace cable(s) immediately if cable(s) insulation becomes damaged.
DO NOT remove fuse or mounting hardware from POSITIVE (+) red battery cable mounting screw.
Part No 1143153
89
11 SETUP/MAINTENANCE
Invacare® Pronto® M91™