Control Panel Rnet - Permobil K Service Manual

Table of Contents

Advertisement

Control Panel R-net
Removal
1. Switch off the main power switch on the control panel.
2. Loosen the control panel cabling from its fixing points. Remember how the cabling is positioned;
this helps when you come to re-attach it. On Corpus 3G, a cover (3) on the the arm rest has to be
removed, it is fitted with two screws.
3. Separate the control panel cabling at the connector on the cabling.
4. Remove the control panel (1). It is held in place with two screws (see illustration). These two screws
also hold the bracket for the ICS control panel (2), if there is one fitted (see illustration).
Assembly
Assemble in the reverse order.
1. Fit the control panel (1) with the two screws (see fig.). Also fit the bracket for the ICS control panel
(2), if the wheelchair is equipped with one, with the same screws (see fig.).
2. Connect the control panel cabling at the connector on the cabling.
3. Re-attach the cabling to its fixing points. On Corpus 3G, refitt the cover (3) on the the arm rest using
the two screws.
1
The control panel ans ICS is attached with two screws.
For this task the following tools are necessary:
1
Allen key 4 mm.
35
35
Control Panel
2

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

C300C300s

Table of Contents