U1273A/U1273AX Handheld Digital Multimeter
Performing Measurements and Tests
Resistance measurements
Set up your multimeter as shown in the figure below to per-
form resistance measurements.
2
Smart measurements:
• Use the Smart (offset compensation) function to measure
resistors affected by DC offset or leakage current.
• Smart removes unexpected DC voltages within the instru-
ment, at the input, or at the circuit being measured, which
will add error to resistance measurements. Press
switch between the bias voltage (BiAS) display or leakage
current (LEAk) display, calculated based on the bias voltage
and corrected resistance value, on the secondary display.
10
4
Smart
1
3
While performing resistance mea-
surements, press
shown on the secondary display to
enable the Smart function.