Resistance Measurements - Keysight u1273a Quick Start Manual

Handheld
Hide thumbs Also See for keysight u1273a:
Table of Contents

Advertisement

U1273A/U1273AX Handheld Digital Multimeter
Performing Measurements and Tests

Resistance measurements

Set up your multimeter as shown in the figure below to per-
form resistance measurements.
2
Smart  measurements:
• Use the Smart  (offset compensation) function to measure
resistors affected by DC offset or leakage current.
• Smart  removes unexpected DC voltages within the instru-
ment, at the input, or at the circuit being measured, which
will add error to resistance measurements. Press
switch between the bias voltage (BiAS) display or leakage
current (LEAk) display, calculated based on the bias voltage
and corrected resistance value, on the secondary display.
10
4
Smart
1
3
While performing resistance mea-
surements, press
shown on the secondary display to
enable the Smart  function.
U1273A/U1273AX Quick Start Guide
until BiAS is
to
Dual
Exit