U1273A/U1273AX Handheld Digital Multimeter
Performing Measurements and Tests
1 Press
setup menu.
2 Browse to Menu 6 > Filter, and press
value.
3 Press
Refer to the table below for the respective firmware ver-
sions.
4 Press
5 Press and hold
to normal operation. Turn the multimeter to DC Voltage
mode to verify that the LPF symbol is turned on.
CAU TI O N
CAU TI O N
LPF during AC measurements:
• Passing the measured signal through a LPF help blocks
unwanted voltages such as electronic noise.
• Use the LPF function to improve measurement on compos-
ite sine waves that are typically generated by inverters and
variable frequency motor drives.
8
for more than 1 second to enter the multimeter's
Setup
Dual
or
Exit
Setup
Version 1.64 or older
on
to save your changes.
Setup
To avoid possible electric shock or personal injury, enable
(
the Filter
voltages. Displayed DC voltages can be influenced by high
frequency AC components and must be filtered to assure
an accurate reading.
to enable the Filter.
Version 1.95 or newer
until the multimeter restarts and returns
)
to verify the presence of hazardous DC
Press
voltage measurements to pass the
measured signal through a low pass
filter.
U1273A/U1273AX Quick Start Guide
to edit the
DC
while performing AC