Connector
Mounting Screw
© National Instruments Corporation
3.
Turn off your SCXI chassis.
4.
Slide the selected module out of the SCXI chassis.
5.
Remove the module cover.
6.
Place one jack screw on the SCXI-1121 as indicated in Figure 2-8.
7.
While holding the jack screw in place, insert the lock washer and then
the nut. Notice that you might need long-nose pliers to insert the
washer and nut.
8.
Tighten the nut by holding it firmly and rotating the jack screw.
9.
Repeat steps 6 through 8 for the second jack screw.
10. Replace the module cover and tighten the grounding screw.
11. Slide the module back in place.
12. Connect the SCXI-1330 to your module connector and secure it by
tightening both mounting screws.
Shell Assembly
Mounting Screw
Jack
Screws
Shell Assembly

Figure 2-8. Assembling and Mounting the SCXI-1330 Connector-and-Shell Assembly

Chapter 2
SCXI-1121 Module
Nut
Lock Washers
Nut
2-23
Configuration and Installation
Grounding Screw
SCXI-1121 User Manual