Chassis 1
Slot
Entry
Number
1
4
2
13
© National Instruments Corporation
Assuming that the modules are cabled and programmed correctly, the
Slot 0 scan lists should be as follows:
Chassis 2
Count
Entry
Number
4
1
6
2
3
Other solutions are possible.
In the section,
3. Programming the Slot 0 Hardscan
chapter, step 3 consists of the following steps:
1.
Select Slot 14 in Chassis 1.
2.
Write
XXXXX001 10000011
3.
Toggle SLOT0SEL*.
4.
Write
XXXXX110 00000101
5.
Select Slot 14 in Chassis 2.
6.
Write
XXXXX110 00000011
7.
Toggle SLOT0SEL*.
8.
Write
XXXXX101 00000001
9.
Toggle SLOT0SEL*.
10. Write
XXXXX110 00000011
11. Select Slot 14 in Chassis 3.
12. Write
XXXXX110 00000101
13. Toggle SLOT0SEL*.
14. Write
XXXXX001 00000000
15. Toggle SLOT0SEL*.
16. Write
XXXXX011 10000010
17. Select Slot 0 in Chassis 0.
Slot
Count
Entry
13
4
11
2
13
4
over MOSI.
over MOSI.
over MOSI.
over MOSI.
over MOSI.
over MOSI.
over MOSI.
over MOSI.
5-17
Chapter 5
Programming
Chassis 3
Slot
Number
Count
1
13
2
3
3
8
Circuitry, earlier in this
SCXI-1121 User Manual
6
1
3

Advertisement

Table of Contents
loading

Table of Contents