Table 2-6. Voltage And Current Mode Excitation Jumper Setup - National Instruments SCXI-1121 User Manual

Four-channel isolated universal transducer module for signal conditioning
Hide thumbs Also See for SCXI-1121:
Table of Contents

Advertisement

Chapter 2
Configuration and Installation
Excitation
Channel
0
1
2
3
SCXI-1121 User Manual

Table 2-6. Voltage and Current Mode Excitation Jumper Setup

Jumpers
W14 and W15
W22 and W23
W34 and W35
W46 and W47
Excitation Level
Each excitation channel of your SCXI-1121 has two different current or
voltage levels. You can set a given channel to one of the following level
modes:
In the Current mode
In the Voltage mode
It is important to notice that you should select the level of excitation
according to the load you are using. Table 2-7 lists the maximum load that
can be driven per channel at each level of excitation for both volt and
current excitation.
Voltage Mode
(Factory Setting)
W14
W15
1
2
3
1
2
3
W22
W23
1
2
3
1
2
3
W34
W35
1
2
3
1
2
3
W46
W47
1
2
3
1
2
3
0.150 or 0.450 mA
3.333 or 10 V
2-10
Current Mode
W14
W15
1
2
3
1
2
3
W22
W23
1
2
3
1
2
3
W34
W35
1
2
3
1
2
3
W46
W47
1
2
3
1
2
3
www.natinst.com

Advertisement

Table of Contents
loading

Table of Contents