Common Questions
Feature
Analog input
Input range
±5 V, ±250 V with SCXI-1327
Isolation
Gains
1, 2, 5, 10, 20, 50, 100, 200, 250, 500,
1000, and 2000 jumper selectable
Filters
4 Hz or 10 kHz jumper selectable
Autozero
Not supported
Calibration
Manually rotateable potentiometers
for one-point offset calibration
Scanning
333 kS/s with consecutive channels
CJC scanning
M TEMP (non-scannable) or
D TEMP (direct channel)
Offset error
±6 µV ± 3 mV/gain
Gain error
±0.2% typ, ±0.6% max
© National Instruments Corporation
This appendix lists common questions related to the use of the SCXI-1125.
The SCXI-1125 is backward compatible with the SCXI-1120, but what
are the major differences between the SCXI-1120 and the SCXI-1125?
Table D-1 compares the major specifications and features of the two
modules. Other specifications and features of the SCXI-1125 are the same
or very similar to the SCXI-1120.
Table D-1. Comparison of the SCXI-1125 with the SCXI-1120
SCXI-1120
8
250 V
rms
±5 V, ±300 V with SCXI-1327,
±1000 VDC with TBX-1316
1, 2, 5, 10, 20, 50, 100, 200, 250, 500,
1000, and 2000 software selectable
4 Hz or 10 kHz software selectable
Software configurable and scannable
Software internal one-point offset
calibration, software external
two-point offset and gain calibration,
and onboard calibration constant
storage in EEPROM
333 kS/s with nonconsecutive and
repeating channels (random scanning)
M TEMP, CJ TEMP (scannable)
0.2 mV/gain typical
±0.03% typ, ±0.08% max
D-1
D
SCXI-1125
8
300 V
rms
SCXI-1125 User Manual