Calibration Of Light Source For Digital Adjustments - Samsung GX-20 Service Manual

Hide thumbs Also See for GX-20:
Table of Contents

Advertisement

2. Calibration of light source for digital adjustments

IMPORTANT
Calibration of light source has to be done prior to the 77012 digital adjustments.
When the adjustment software, the light source or the photometry standard lens for adjustments are
replaced, calibration must be done. (Calibration process is identical to the 76832 (GX10).)
Preparation Equipment
Master body for 76830(K10D) light source calibration
77012 digital adjustments software (M-Test)
PC (Windows XP, USB port standard device)
Light box (LB-3300, Light source A)
Photometric standard lens for adjustments and F8 set ring
Use a lens with an ID no. identical to the lens ID no. as shown on the CD-R (adjustments software).
USB cable
A C adaptor
Material for blocking light (dark curtain)
Color thermometer for photography (for light source adjustments)
LV checker (for checking LV values)
1. Computer Set-up
2 . Setting the computer
(Use the digital adjustment software.)
in the
P reparation
- This Document can not be used without Samsung's authorization -
section has to be completed.
. ADJUSTMENT
73