Calibration/ Adjustment

Always calibrate/ adjust the balance/ scale after setting it up in a new location.
Available Features
Calibration/ adjustment can only be performed when
there is no load on the balance/ scale,
the balance/ scale is tared,
the internal signal is stable.
If these conditions are not met, an error message is displayed.
The weight required for calibration/ adjustment is displayed (see Accessories for
calibration weights; for model GC503 see Equipment Supplied ).
lite
On Basic
balances, calibration/ adjustment can be performed
with any of three weight units: g, kg* * , or lb (menu code 1. 4. x)
Factory setting: Grams (menu code 1. 4. 1)
Example
Calibrate the balance/ scale
Menu code settings: Factory-set codes
Step
1. T urn on the balance/ scale
2. T are the balance/ scale
3. Begin calibration
Calibration weight is displayed
without weight unit (here: 1000 g)
4. Place the indicated calibration
weight on the balance/ scale
After calibration, the calibration
weight is displayed with wt. unit
5. R emove the calibration weight
*
Calibration technically means to determine the difference between the balance/ scale readout and the
actual weight on the pan to determine the accuracy.
Adjustment means to bring a balance/ scale into the state of accuracy required for its use.
Therefore, calibration, as used in this manual, actually means adjustment.
* * = not on models with a readability of 0.1 mg
*
Key (or instruction) Display/ Printout
e
t
>2 sec.
t
0.0
g
+1000.0
1000.0
+1000.0
g
0.0
g
9