Initialization; Resolution Setting - Jadever JWE Series Service Manual

Hide thumbs Also See for JWE Series:
Table of Contents

Advertisement

Linear Calibration
1)Press and hold key
release the keys.
MODE
2) Press key
3) Wait till
TARE
key
E.g., for JWE-3K ,1/3 of full load is 1kg.
4) Wait till
TARE
E.g., for JWE-3K, 2/3 of full load is 2kg.
5) Wait till
JWE-3K, full load is 3kg.
6) The calibration procedure ends up with a symbol of "PASS".
7) Remove all the weights, and then press key

4-3 Initialization

1. Press and hold key
"SET"
MODE
2. Press key
3. Press and hold key
ZERO
4. Press
to go back to weighing mode

4-4 Resolution setting

1. Press and hold key
MC/SET
2. Press key
TARE
3. Press
to select 1 or 2. (1 for 1/15000; 2 for 1/30000)
ZERO
4. Press
to save and return
JADEVER SCALE CO., LTD
TARE
while powering on the scale. The screen displays
to enter the zero point calibration mode.
appears, put weights of 1/3 of full load on the weighing pan and press down
appears, and then put weights of 2/3 of full load on pan and press down key
appears, then put weight of full load on and press
TARE
ZERO
and
, and the screen will display "FACT"
M+/MR
TARE
and
M+/MR
while powering on the scale
MODE
TARE
,
,
JWE service manual
ZERO+
to return to weighing mode.
while powering on the scale. The screen will display
to start linear calibration
one by one. And the screen will display "SENC"
TARE
E.g., for
, then
13