Electrical Specifications - Iso-Tech IDM 507 Instruction Manual

Process multimeter
Table of Contents

Advertisement

Electrical Specifications

˙ Accuracy is ± (% reading + number of digits)
˙ Ambient temperature: 23°C ± 5°C (< 80% RH)
˙ For the best measurements, with REL Δ function to
compensate for offsets.
1. Voltage
Function
Range
50.000mV
500.00mV
[1]
AC
5.0000V
50.000V
500.00V
[2]
1000.0V
50.000mV
500.00mV
5.0000V
DC
50.000V
500.00V
1000.0V
[1] Below 5% of AC range, add 20d to accuracy.
[2] The bandwidth of range is 40Hz to 1kHz.
Input Protection : 1000V
DC
Input Impedance : 10MΩ, < 100pF
Bandwidth : 40Hz to 10kHz
Minimum Resolution : 1µV in the 50mV range
CMRR / NMRR (Common / Normal Mode Rejection Ratio) :
V
: CMRR > 60dB at DC, 50Hz / 60Hz
AC
V
: CMRR > 100dB at DC, 50Hz / 60Hz
DC
NMRR > 50dB at DC, 50Hz / 60Hz
AC Conversion Type :
AC conversions are ac-coupled, true rms responding,
calibrated to the sine wave input.
For non-sine wave add the following Crest Factor corrections :
For Crest Factor of 1.4 to 2.0, add 1.0% to AC accuracy.
For Crest Factor of 2.0 to 2.5, add 2.5% to AC accuracy.
For Crest Factor of 2.5 to 3.0, add 4.0% to AC accuracy.
AC+DC Accuracy : AC accuracy + DC accuracy + 1.0%
HFR Accuracy : AC accuracy + 1.0% for 40Hz to 400Hz
The Cut-Off Frequency of HFR : 800Hz (-3dB point)
Attenuation Characteristic of HFR : Approx -24dB
Accuracy
Sine Wave:
± (0.7% + 20d) for 40Hz to 70Hz
± (1.5% + 40d) for 71Hz to 10kHz
Sine Wave:
± (0.5% + 20d) for 40Hz to 70Hz
± (1.5% + 40d) for 71Hz to 1000Hz
± (3.0% + 80d) for 1001Hz to 10kHz
± (0.05% + 30d)
± (0.05% + 5d)
or 1000V
rms
AC