Electrical Specifications

Accuracy is ±(% reading + number of digits) at 23°C ± 5°C < 80%RH.
(1) Voltage
IDM98-III/IDM99-III
Range
Resolution
60.00mV
0.01mV
DC
600.0mV
0.1mV
DC
6.000V
0.001V
DC
60.00V
0.01V
DC
600.0V
0.1V
DC
1000V
1V
DC
60.00mV
0.01mV
AC
600.0mV
0.1mV
AC
6.000V
0.001V
AC
60.00V
0.01V
AC
600.0V
0.1V
AC
1000V
1V
AC
Input Protection : 1000VDC or 1000VAC rms
AC Frequency Response : 50Hz ~ 1KHz
Input Impedance : 10MΩ, <100pF
AC Conversion Type :
AC conversions are ac-coupled, true rms responding, calibrated to the
sine wave input.
For non-sine wave add the following Crest Factor corrections:
For Crest Factor of 1.4 to 2.0, add 1.0% to accuracy.
For Crest Factor of 2.0 to 2.5, add 2.5% to accuracy.
For Crest Factor of 2.5 to 3.0, add 4.0% to accuracy.
18
IDM 98III / 99III English
IDM98-III Accuracy
IDM99-III Accuracy
± (0.08% + 10d)
± (0.1% + 2d)
± (0.08% + 2d)
± (0.09% + 2d)
± (0.08% + 2d)
± (0.09% + 2d)
± (0.08% + 2d)
± (0.09% + 2d)
± (0.08% + 2d)
± (0.09% + 2d)
± (0.08% + 2d)
± (1.20% + 5d)
± (1.5% + 5d)
± (1.20% + 5d)
± (1.0% + 3d)
± (0.80% + 5d)
± (1.0% + 3d)
± (0.80% + 5d)
± (1.0% + 3d)
± (0.80% + 5d)
± (1.0% + 3d)
± (0.80% + 5d)
IDM 98III / 99III English
(2) Auto-V
IDM98-III/IDM99-III
Range
Resolution
600.0V
0.1V
DC
1000V
1V
DC
600.0V
0.1V
AC
1000V
1V
AC
Input Protection : 1000V
or 1000V
DC
AC Frequency Response : 50Hz ~ 1KHz
Input Impedance : Approx. 3kΩ.
AC Conversion Type :
Conversion type and additional specification are same as voltage
function.
(3) Current
IDM98-III/IDM99-III
Range
Resolution
60.00mA
0.01mA
DC
600.0mA
0.1mA
DC
6.000A
0.001A
DC
10.00A
0.01A
DC
60.00mA
0.01mA
AC
600.0mA
0.1mA
AC
6.000A
0.001A
AC
10.00A
0.01A
AC
Input Protection : Protected by high energy fuses.
440mA, 1000V AC/DC 10KA Fast acting fuse for mA input. 11A, 1000V
AC/DC 20KA Fast acting fuse for A input
Max. Measuring Time : 3 minutes for "A" input, 10 minutes for "mA"
input.
Rest time 20 minutes min.
AC Conversion Type : Conversion type and additional specification are
same
as voltage function.
AC Frequency Response : 50Hz ~ 1KHz
IDM98-III Accuracy
IDM99-III Accuracy
± (1.0% + 3d)
± (0.80% + 3d)
rms
AC
IDM98-III Accuracy
IDM99-III Accuracy
± (1.0% + 3d)
± (0.8% + 3d)
± (1.0% + 3d)
± (0.8% + 3d)
± (1.0% + 3d)
± (0.8% + 3d)
± (1.0% + 3d)
± (0.8% + 3d)
± (1.5% + 3d)
± (1.2% + 3d)
± (1.5% + 3d)
± (1.2% + 3d)
± (1.5% + 3d)
± (1.2% + 3d)
± (1.5% + 3d)
± (1.2% + 3d)