Dc Current Output - Iso-Tech IDM 507 Instruction Manual

Process multimeter
Table of Contents

Advertisement

DC Current Output

˙ To use the DC current output function, turn the rotary in output
position
(Adjustable DC output or Auto DC
˙ The DC current output function has both modes:
& Simulate Mode
˙ The output mode has both types: 0-20mA & 4-20mA. That is
selectable. When power on, press the RANGE button to
select. And the selection was set to default.
Adjustable DC Current Output
˙ To use the adjustable DC current output function, turn
the rotary in
adjustable
output position (
˙ In this function, you can adjust the DC current output.
˙ %STEP: 0% / 25% / 50% / 75% / 100% / 120% / 125%
˙ Fast Setup: 0% / 50% / 100%
˙ Fine Setup : Minimum resolution 1uA, 0mA to 24mA
% STEP
0-20mA Mode
0%
25%
50%
75%
100%
120%
125%
Auto DC Current Output
˙ To use the auto DC current output function, turn the rotary in
auto
output position (
˙ In this function, you can press the BLUE button to
select 4 kinds of the auto DC current output.
˙ Press the HOLD button to pause / continue the output.
Mode
Type
Linear
Linear
25% Step
25% Step
output).
Source Mode
)
4-20mA Mode
0mA
4mA
5mA
8mA
10mA
12mA
15mA
16mA
20mA
20mA
24mA
N/A
N/A
24mA
)
Action
0% to 100% to 0% per 40 sec
0% to 100% to 0% per 20 sec
0% to 100% to 0%, a step per 15 sec
0% to 100% to 0%, a step per 5 sec
9

Advertisement

Table of Contents
loading

Table of Contents