Infopia Element Neo Instructions For Use Manual page 13

Table of Contents

Advertisement

CAUTION
1. If the blood smears or runs, do not use such blood sample. Dry the area and gently
squeeze another drop or puncture a new area of the finger.
2. Do not share reusable lancing device with anyone including other family members.
4. Put the lancing tip back and twist it tightly.
5. To adjust the depth setting:
Use strength level 1-2 for soft skin, 3 for
average skin, 4-5 for thick or calloused skin.
6. Pull the end of the lancet device back.
7. To prick your finger, push the center button.
• 24 •
Step 3
• Applying Blood
• Correct
Completely Filled
• Incorrect
Poorly filled
®
Element
NEO Test Meter only requires 0.5 ul sample volume.
• Stick the edge of the test strip to the drop of blood.
• The blood will automatically be 'drawn' into the test strip channel.
• When your sample blood is enough, your meter will automatically count down.
• Apply your blood to the glucose test strip and do not take off your finger until you get the
beep sound. The test will automatically start.
• Please respect recommended quantity of blood sample. Too much blood may cause
malfunction of your meter by contaminating it.
CAUTION
1. If the countdown does not start, do not add more blood to the glucose test strip.
Discard the glucose test strip and re-start testing.
2. If you do not conduct the test within 3 minutes, the glucose test meter will automatically
power off to save battery life. In this case, test procedure should start again from the
beginning.
3. You may get an inaccurate result if the blood sample is not completely filled.
Correct
Incorrect
• 25 •