Animas OneTouch Ping Owner's Booklet page 80

Glucose management system
Hide thumbs Also See for OneTouch Ping:
Table of Contents

Advertisement

CHAPTER 11 - USING ADVANCED FEATURES
5.
The ezBG Total screen is displayed with a suggested bolus amount. The bolus amount field is
highlighted and flashing to indicate you can enter the total amount as shown or adjust as needed.
Enter your bolus amount. Press
ezBG Total
BG
+
2.81U
IOB
- -.- - U
Total =
2.80U
0.00U
0.00U
Go
Main Menu
ezBG Total
NOTE: If the IOB feature is activated, your pump will calculate a reduced
BG
+
2.81U
BG Bolus amount for your review.
IOB
-
1.00U
Total =
1.80U
1.80U
Go
Main Menu
NOTE: If you enter a BG amount below 70 mg/dL or above 250 mg/dL, your pump will alert you that
you've entered an out of range BG. To confirm the Alert, press
recommended by your health care team.
Alert
LOW BG
Treat low BG.
No bolus
recommended.
Monitor BG.
Confirm
to deliver.
ezBG Total
BG
+
2.81U
IOB
- -.- - U
Total =
2.80U
2.80U
Go
Main Menu
Alert
HIGH BG
Treat high
BG.
Check site.
Check
ketones.
Monitor BG.
Confirm
66
DELIVERING
2.80
U
Press any
function button
to cancel.
. Treat the out of range BG as