CHAPTER 2 - ESTABLISHING COMMUNICATION BETWEEN YOUR METER REMOTE AND YOUR PUMP
3
Confirm the unpairing
Press
on your meter remote to confirm the Notification on your meter
remote.
A second Notification screen will appear on your meter remote display to let
you know that your meter remote and pump are now unpaired. Press
your meter remote display to confirm the Notification. You will go to the
Meter Home screen on your meter remote display (see Chapter 3 in Section
III, pages 187–188).
NOTE: There is no separate unpairing procedure on your pump. Your pump remains ready to
re-establish an RF link with the last paired meter remote, or to pair with a new meter remote.
Reactivating/deactivating the RF feature on your meter remote
NOTE: The "RF on/off" menu option on the RF set-up screen on your meter remote is only available if
your meter remote and pump are paired.
Deactivating RF communication on your meter remote
There are times when you might want or need to deactivate the RF feature on your meter remote. One
situation is when you are on an airplane. Follow these instructions for deactivating the RF feature on
your meter remote if it is activated.
1
Go to RF on/off on your meter remote display
Highlight "RF on/off" on the RF Set-up screen on your meter remote display
and then press
.
2
Deactivate the RF feature on your meter remote
You will be reminded on your meter remote display that the RF feature is
activated. To deactivate the RF feature, highlight "RF off" on your meter
remote display and press
and pump will be stopped.
3
Wait for confirmation that the RF feature on your meter remote is
deactivated
A Notification screen will appear on your meter remote display to remind
you that the RF feature is deactivated on your meter remote.
Press
on your meter remote to confirm the Notification. You will go to the
Meter Home screen on your meter remote display (see Chapter 3 in Section
III, pages 187–188).
. All communication between your meter remote
182
Notification
Pump unpaired.
on
Confirm
Meter remote display
RF Set-up
RF on/off
RF Channel
Pairing
RF Test
Meter remote display
RF on/off
The RF is active. Caution!
All communication to
connected RF devices will
be stopped with RF off!
RF off
Cancel
Meter remote display
Notification
RF communication
deactivated.