CHAPTER 10 - ADVANCED FEATURES / SETUP AND ACTIVATION
To return contrast setting to original factory default, press the
time. When the word "Contrast" is displayed on the screen, press any button to return to the default
contrast setting.
NOTE: When viewing your pump display in bright sunlight, it is recommended you shade the screen or
move to a shady area for best visibility.
Setup Advanced Screen 5 – Auto-OFF Feature
This screen allows you to set your pump to automatically suspend basal delivery and sound an alarm if
no buttons are pressed in a user-selected number of hours. This feature can be used as a safeguard in
case the user is unconscious.
1.
Scroll to the desired field.
2.
Press
to change to flashing cursor for Edit mode.
3.
Use the
/
buttons to change to desired setting.
4.
Press
when setting is made.
To move to the next Setup Advanced screen, scroll to the bottom of the screen
and highlight Next. Press
To return to the Home screen, scroll to "Home" and press
Setup Advanced Screen 6 – Low Cartridge Warning Setting and
Occlusion Sensitivity Setting
This screen allows you to:
Set your low cartridge warning to alert you at 10, 20, 30, 40 or 50 units remaining
Set your occlusion detection sensitivity to High (more sensitive) or Low (less sensitive)
1.
Use the
/
buttons to scroll to the desired field.
2.
Press
to change to flashing cursor for Edit mode.
3.
Use the
/
buttons to change to desired setting.
4.
Press
when setting is made.
To move to the next Setup Advanced screen, scroll to the bottom of the
screen and highlight "Next". Press
To return to the Home screen, scroll to "Home" and press
.
.
52
button and
button at the same
SETUP ADV 5
Enable
suspend in 12 Hr
Home
.
SETUP ADV 6
Low Cartridge
Warning
Occlusion
Sensitivity
Home
.
AUTO OFF
OFF
When enabled,
if no buttons
are pressed.
Next
20U
L
Next