CHAPTER 1 - IMPORTANT INFORMATION
Medical Procedures and Equipment
Procedure
Pacemaker/
Automatic Implantable
Cardioverter-
Defibrillator (AICD)
EKG
Cardiac
Catheterization
Nuclear Stress Test
Colonoscopy
Laser Surgery
General Anesthesia
For procedures not included above or on the chart on the following page, call (in advance of your
procedure) 1 877 937-7867 in the U.S., Canada and Puerto Rico.
Patient is Pumper
Disconnect pump and leave outside
room during insertion of device and
reprogramming. Infusion set can
remain in place.
No need to disconnect.
Disconnect pump and leave outside
room during procedure.
Disconnect pump and leave outside
room during scan. Can remain
connected during injection of
radioisotope.
No need to disconnect.
Pump and infusion set can be worn;
however some lasers can create
interference and cause pump to alarm.
Determination based on what medical
equipment is being used in the
procedure.
4
Health Care
Professional/Technician
is Pumper
Follow normal safety practices and
proceed to the designated safe
area while each X-ray is taken and
during reprogramming.
No need to disconnect.
Follow normal safety practices and
proceed to the designated safe
area while each X-ray is taken.
Follow normal safety practices and
proceed to the designated safe
area while each X-ray is taken.
No need to disconnect.
Pump and infusion set can be
worn; however some lasers can
create interference and cause
pump to alarm.
Determination based on what
medical equipment is being used
in the procedure.