CHAPTER 4 - GETTING YOUR PUMP READY
Priming your Pump and Infusion Set
NOTE: As each step is completed, the check box on the ezPrime menu will be shaded.
WARNING: Never prime the tubing or tighten the cartridge cap while infusion set is
connected to your body. Priming the tubing or tightening the cartridge cap while the infusion
set is connected to your body can result in serious injury or death.
1.
Make sure you are disconnected from your pump.
2.
MAIN MENU
Bolus
Suspnd/Resum
History
Basal
Setup
Prime/Rewind
Status
Home
3.
On the ezPrime menu, "Rewind" is highlighted.
Press
. The REWIND MOTOR screen is displayed.
4.
Scroll up to "Go Rewind". Press
will rewind the piston rod.
NOTE: Before starting the rewind, your pump will
vibrate as it performs a self test.
5.
When the rewind action is complete, the REWIND COMPLETE screen is displayed. Your pump will
beep once to let you know the rewind is complete.
REWIND
ACTIVE
NOTE: If using a partially filled cartridge, you can select Stop during the Rewind
Position:
Active function to stop the rewind at the position desired. After every third
160U
rewind, your pump is required to do a Full Rewind and will not offer the option
of selecting the "Stop" position. A Full Rewind is always required when a
------
------
------
<<
<<
<<
battery is inserted.
Stop
From the MAIN MENU, select "Prime/Rewind".
. Your pump
22
ezPrime
o
Rewind
o
Load Cart
o
Prime
o
Fill Cannula
Main Menu
REWIND MOTOR
Disconnect
infusion set
from your body!
Go Rewind
Cancel
REWIND MOTOR
Disconnect
infusion set
from your body!
Go Rewind
Cancel
REWIND
ACTIVE
Please wait.
Full rewind
required.
------
------
------
<<
<<
<<