Table 20 Summary Of Pressure Transmitter Configuration Parameters; Overview - Honeywell ST 3000 User Manual

Smart transmitter release 300 and smart field communicator model sts103
Hide thumbs Also See for ST 3000:
Table of Contents

Advertisement

6.2

Overview,

What to configure
ATTENTION
Table 20
Summary of Pressure Transmitter Configuration Parameters
Configuration Data
Transmitter Tag Number
Damping Time Constant
Type of Output Conformity
Unit of Measurement
LRV (Lower Range Value)
(Process input for 4 mAdc
(0%) output)
URV (Upper Range Value)
(Process input for 20 mAdc
(100%) output)
February 2012
ST 3000 Release 300 and SFC Model STS103 User's Manual
Continued
Table 20 summarizes the parameters that are included in the
configuration database for an ST 3000 pressure transmitter in either the
analog or DE mode of operation.
Be aware that configuration data for the transmitter as well as for the
Local Smart Meter is stored in a non-volatile memory on the
transmitter's PWA and make up the transmitter's configuration
database. Therefore, the transmitter and meter configuration are lost if
the PWA is replaced. Performing a save and restore function using the
SFC will preserve the transmitter's configuration database. See Section
8.5 for the steps to perform save and restore functions using the SFC.
Since the SFC is compatible with other Honeywell Smartline
transmitters, be sure all configuration data applies to a pressure
transmitter.
Up to eight characters
Any one of these value selections in seconds:
0.00
0.5
0.2
1.0
0.3
2.0
LIN (Linear)
SQRT (Square Root)
ATTENTION Note that ST 3000 transmitters with inches of water
ranges are factory calibrated using pressure referenced to a temperature
of 39.2°F (4°C).
Pressure readings can be displayed in any one of these pre-programmed
engineering units:
"H2O_39F
PSI
mmHg_0C
KPa
"H2O_68F
ATM
Key in desired value through SFC keyboard or set LRV to applied
pressure.
Key in desired value through SFC keyboard or set URV to applied
pressure.
Setting or Selection
4.0
32.0
8.0
16.0
MPa
bar
KG/cm^2
mbar
G/cm^2
inHg_32F
"H2O_60F
mmH2O_4C
mH2O_4C
Continued on next page
63

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents