Verifying Configuration Data; Table 3 Verifying Transmitter's Configuration Data (Using The Sfc) - Honeywell ST 3000 Installation Manual

Smart transmitter
Hide thumbs Also See for ST 3000:
Table of Contents

Advertisement

2.3

Verifying Configuration Data

Procedure
ATTENTION
Table 3
Verifying Transmitter's Configuration Data (Using the SFC)
Step
Press Key
1
DE READ
A
ID
NON-VOL
ENTER
(Yes)
2
C
DAMP
10
Use the procedure in Table 3 to display all the basic transmitter database
parameters to be sure they are correct. Note that the values/selections
shown in displays are for example purposes only.
• This procedure assumes that you have established communications
with the transmitter as outlined in Table 2.
• If any parameter is not set to the correct value/selection or your
transmitter was not configured, you can do so now. Refer to the
Configuration Section in the ST 3000 Smart Transmitter, Release 300
and SFC Smart Field Communicator Model STS 103 User's Manual,
34-ST-25-14 for details.
Read Display or Action
T A
G
N
O
.
T R I P S
S E C U R E D ?
T A
G
N
O
.
S F C
W O R K I N G . .
L
I N
D P
T
S
D A M P
1
S
Ø .
3
S
ST 3000 Release 300 Installation Guide
This prompt only appears for
transmitters in analog mode.
?
This is only required for transmitters
in analog mode.
.
A G
N O .
Transmitter's assigned tag number
P T
1 Ø Ø 1
"SPT 1001" appears in bottom row.
Present damping time setting.
P T
1 Ø Ø 1
E C
O
N
D
S
Description
Continued on next page
2/05

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents