3.7 Changing an existing configuration

When a configuration is set up, you can add or remove displays
without running the Configuration Setup Wizard.
Note:
After changing anything to the configuration, you need to run the Q/A
tasks again. If not, the task results will appear Unknown.

Adding a display

If the main window displays an imaging board head without an added
display (represented by a square without display icon), you can add a
display to the head.
To add a display:
1 Right-click on the imaging board head icon and select Add Display...
from the drop-down menu.
Using MediCal Pro
Browse and select your own test pattern. MediCal Pro
supports bmp, jpg and gif files.
This pattern is a universal pattern with which you can
check geometry, focus, white luminance, bandwidth
and display curve, according to the AAPM qualifica-
tions.
This pattern can be used to check if the white and
black levels are uniform all over the screen. There-
fore, put the sensor on the different patches on the
screen, and check the Measure window to see if the
measured value is approximately the same all over
the screen.
This pattern shows three different white-to-black gra-
dients. You can use it to check the grayscale perfor-
mance and bandwidth.