As a result, possible purity errors in the image (so-called hot spots) are
removed. The display automatically degausses at switch-on.

3.14 Special functions for displays in mammography systems

These functions are available for Barco displays of type MFGD 5621 HD
only.
To get access to the special functions:
1 In the Configuration section in the main window, right-click on the icon
of the display you wish to view the properties of.
2 From the drop-down menu that appears, select Properties...
3 Select the Display tab in the display device properties dialog.
4 Click on Details...
5 The display details window appears.

Per Pixel Uniformity

For mammography viewing applications we recommend switching Per
Pixel Uniformity (PPU) on.
When switched on, the Per Pixel Uniformity system is activated, adjusting
the luminance output at the pixel level and eliminating screen noise.
If you wish to activate PPU, check the Per Pixel Uniformity check box. To
de-activate PPU, uncheck the Per Pixel Uniformity check box.

Show Pixel Defects

Click this button to get a graphical overview of pixel artifacts and
microblemishes on the LCD panel.
Using MediCal Pro
102