Calibrating Barco Displays With Built-In I-Guard Sensor; Calibrating Barco Displays With External Sensor - Barco MediCal Pro Installation & User Manual

Hide thumbs Also See for MediCal Pro:
Table of Contents

Advertisement

You can also calibrate a display by right-clicking on its icon in the main
window, and selecting Calibrate... from the drop-down menu.

Calibrating Barco displays with built-in I-Guard sensor

The Barco Coronis flat panel displays have a built-in optical sensor that
communicates with MediCal Pro.
The calibration runs completely automatically.
When the calibration is finished successfully, it displays the following
message.
Click on OK to close the window.

Calibrating Barco displays with external sensor

Some Barco flat panel displays, like Nio and Nio Fusion, do not have an
internal sensor and must be calibrated with an external sensor.
Proceed as follows:
1 For certain types of sensors, you may be asked to calibrate the sensor
first.
Make sure the sensor is completely covered (so that no light can enter
the sensor) and click on OK.
2 Put the sensor, with Ambient Light Shield if appropriate, on the image
on the screen.
Using MediCal Pro
69