Percent Oxygen Analyzer
Probes of 1000 mm (40") normally have sufficient length for any
installation. Customers requiring probes longer than 1500 mm (59") are
supplied with a flow guide tube that uses the flue velocity to pull flue
gas through a filter at the sensing tip and exhaust it near the flue wall.
2.2.1 Probe Configuration
The probe can be configured in a variety of ways to suit the
analytical environment. Figure 2-1 shows the most common
configurable items of the probe. Not indicated is the actual
thermocouple sensing element, however, see Table 2-1. You can choose:
1. Probe insertion length (from process end of mounting thread to
probe sensing tip). See Figure 2-2.
2. Mounting thread (process connection), BST or NPT (for size of
thread refer to specifications).
3. Lagging extension length, if required.
Table 2-1 indicates typical probe options.
Table 2-1: Probe Options
Probe Length
250 mm (10")
350 mm (14")
500 mm (20")
750 mm (30")
1000 mm (40")
Thermocouple
Type K max 900°C (1560°F)
Figure 2-2: Model 9060L Probe
Teledyne Analytical Instruments
Operational Theory
Sheath
316 SS max 850°C (1560°F)
Inconel
The Inconel option has all
inconel wetted parts except for
the ceramic sensor and viton
'o' rings
Thread
BST or NPT
17