Page 1
OPERATING INSTRUCTIONS FOR Model 9060L Percent Oxygen Analyzer P/N M77713 ECO: DANGER Toxic gases and or flammable liquids may be present in this monitoring system. Personal protective equipment may be required when servicing this instrument. Hazardous voltages exist on certain components internally which may persist for a time even after the power is turned off and disconnected.
Page 2
Teledyne Instruments/ Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749- 1580.
Page 3
Percent Oxygen Analyzer Specific Model Information Instrument Serial Number: _______________________ Instrument Range: _______________ Calibrated for: _______________ Background Gas: _______________ Zero Gas: _______________ Span Gas: _______________ Teledyne Analytical Instruments...
Page 4
Model 9060L Safety Messages Your safety and the safety of others is very important. We have provided many important safety messages in this manual. Please read these messages carefully. A safety message alerts you to potential hazards that could hurt you or others.
Page 5
Manuals do get lost. Additional manuals can be obtained from TI/AI at the address given in the Appendix. Some of our manuals are available in electronic form via the internet. Please visit our website at: www.teledyne-ai.com. Teledyne Analytical Instruments...
List of Figures Figure 1-1: Front Panel ..............12 Figure 1-2 Rear Panel ..............14 Figure 2-1: Model 9060L Oxygen Probe ........16 Figure 2-2: Model 9060L Probe............. 17 Figure 2-3: Internal Structure of the Zirconia Sensor and Probe ... 18 Figure 2-3: Signal Processing Block Diagram .......
Percent Oxygen Analyzer List of Tables Table 2-1: Probe Options .............. 17 Teledyne Analytical Instruments...
Page 10
Since the use of this instrument is beyond the control of Teledyne Instruments/ Analytical Instruments, referred as TI/AI, no responsibility by TI/AI, its affiliates, and agents for damage or injury from misuse or neglect of this equipment is implied or assumed.
1.1 Overview The Teledyne Instruments/ Analytical Instruments (TI/AI) Model 9060L is a microprocessor-based percent oxygen analyzer for real-time measurement of the percent of oxygen in inert gases, or in a wide variety of gas mixtures. It features simple operation, fast response, and a compact, rugged construction.
Introduction Model 9060L • Two concentration alarms with adjustable setpoints. • A failure alarm. • Three analog outputs: two for measurement (0–10VDC, and negative ground 4–20 mA DC) and one for range identification (0-10VDC). • Compact and rugged control unit with flush-panel case.
1.4 Rear Panel Description The rear panel contains the electrical input and output connectors. The connectors are described briefly here and in detail in the Installation chapter of this manual. Teledyne Analytical Instruments...
Introduction Model 9060L Figure 1-2 Rear Panel • Power Connection 110 VAC at 50/60 Hz. The connector housing includes the fuse holder and the power switch. (220VAC is optional via setting an internal switch). Fuse Holder: Replacing the fuse is described in Chapter 5, Maintenance.
• Reduced probe breakage due to greater resistance to thermal shock and mechanical damage during installation and startup. The 9060L probes or sensors are simple to install and maintain. All models provide direct measurement of oxygen level. Figure 2-1 shows a typical oxygen probe.
Operational Theory Model 9060L All Teledyne oxygen probe or sensors are designed and manufactured to exacting standards of performance and reliability and are the result of extensive research and development. Teledyne Analytical Instruments provides worldwide application and after sales support for its oxygen probes, sensors and analyzers.
BST or NPT 350 mm (14”) Inconel 500 mm (20”) The Inconel option has all 750 mm (30”) inconel wetted parts except for the ceramic sensor and viton 1000 mm (40”) ‘o’ rings Figure 2-2: Model 9060L Probe Teledyne Analytical Instruments...
Figure 2-3. Figure 2-3: Internal Structure of the Zirconia Sensor and Probe Heater control in the Model 9060L Oxygen Analyzer consists of a time proportioning temperature controller and solid state relay so that the thermocouple junction is controller to over 700°C (1300°F).
Outside (ATM) 0.21 ⎝ ⎠ The 9060L transmitter solves this equation which is valid above 650°C (1200°F). Either the probe heater or the process maintains the sensor temperature at this level. 2.2.3 Heater CAUTION: DANGEROUS VOLTAGE (120 VAC) EXISTS WITHIN...
For high and low pressure applications where the pressure at the point of measurement is significantly above or below atmospheric pressure, compensation must be applied. The Model 9060L does not incorporate a pressure sensor thus no compensation is performed. This factor must be considered when reading concentration from the display under high or low pressure monitoring conditions.
The output from the sensor amplifier is sent to an analog to digital converter (ADC), and the resulting digital concentration signal is passed to the micro-controller. The ADC has a second channel which monitors the temperature of the probe. Teledyne Analytical Instruments...
Page 22
Operational Theory Model 9060L The digital concentration signal along with input from the front panel buttons (KEYBOARD) is processed by the micro-controller, and appropriate output signals are directed to the display and alarm relays. The same digital information is also sent to a 12-bit digital to analog converter (DAC) that produces the 0-10 VDC analog concentration signal and the 0-10 VDC analog range ID output.
3.2 Control Unit Installation The 9060L control unit is designed to be panel mounted in a general purpose, indoor area, away from moisture and the elements. The unit should be installed at viewing level in a sheltered area.
Page 24
If the probe is to be installed on a bend in the flue, it is best to locate it on the outer circumference of the bend to avoid dead pockets of flue gas flow. While the standard 9060L probe with a ‘U’ length of 250 mm (10”) will suit most low temperature flue applications, it is occasionally necessary to have a longer probe with the sensing tip in the center of the flue gas stream.
Figure 3-1: Oxygen Probe Mounting 3.4 Electrical Connections Figure 3-1 shows the Model 9060L rear panel. For detailed pinouts, see the wiring/interconnection drawings in the Drawings section at the rear of this manual. 3.4.1 Primary Input Power The power cord receptacle, fuse block and Power switch are located in the same assembly.
Installation Model 9060L Figure 3-2: Model 9060L Rear panel 3.4.2 Analog Outputs There are three DC output signal connectors with screw terminals on the panel. There are two wires per output with the polarity noted. See Figure 3-3. The outputs are: 0–10V % Range:...
Actuates when the output of the sensor falls below a Sensor Fail factory programmed acceptable level or the thermocouple is open. CAUTION: THERE COULD BE HAZARDOUS VOLTAGE AT THE ALARMS TERMINALS, EVEN WHEN POWER IS REMOVED FROM THE INSTRUMENT. Teledyne Analytical Instruments...
Figure 3-4 . The heater connector is on the right-hand corner of the rear panel. Six foot long cables are provided for the remote probe and the heater. Their Teledyne Analytical Instruments part numbers are listed on the spare parts list located in the appendix.
• Checked the probe and welded probe socket for leaks Once the above checks have been made, you can connect the power source. Wait approximately 30 minutes for warm up, two hours wait is best. After that, the instrument is ready for operation. Teledyne Analytical Instruments...
Page 30
Installation Model 9060L Teledyne Analytical Instruments...
Analyzing for the percent oxygen level in the gas passing through the probe and sensor is the default mode of operation. As long as no front panel buttons are being pressed the analyzer is analyzing. Figure 4-1: Front Panel Controls and Indicators Teledyne Analytical Instruments...
Operation Model 9060L 4.2 Using the Function and Data Entry Buttons When no buttons on the analyzer are being pressed, the instrument is in the analyze mode. It is monitoring the percent of oxygen in the sample gas that is flowing through the remote probe.
2. To change the setting at which the alarm will be actuated, press the SET ALARM 1 function button once more. The alarm setpoint will flash on the LED display. Press either the Up or Down keys to raise or lower the displayed value, as required, Teledyne Analytical Instruments...
650°C, or the thermocouple is sensed as open. 4.5 Selecting a Fixed Range or Autoranging The Model 9060L can operate in fixed high, fixed low, or autoranging mode. To change modes: 1. Press and then release the SET HI RANGE and the SET LO RANGE buttons simultaneously.
Set the alarm setpoints and the full scale ranges to the desired values. Figure 4-2: Calibration Port on Probe Head Span Procedure: 1. Press the SPAN button once. Note: The numeric LED starts flashing and will continue to flash Teledyne Analytical Instruments...
Page 36
Operation Model 9060L for five minutes before timing-out. During this five minute interval, the LED will continue to track gas through the analyzer. 2. Flow span gas of a known oxygen concentration (between 1 and 16% oxygen) into the analyzer. The analyzer will begin to track the span gas.
The probe temperature can be displayed momentarily and adjusted if needed. While the instrument is in normal mode, displaying oxygen concentration, press the UP and DOWN arrow keys simultaneously. The red LED display will show the probe temperature in degrees centigrade for five seconds. Teledyne Analytical Instruments...
Operation Model 9060L Within the time the temperature is being displayed, one can use the UP and DOWN arrow keys to adjust the temperature, thus recalibrating the temperature readout. WARNING: TEMPERATURE READING IS CALIBRATED AT THE FACTORY. DO NOT ATTEMPT RECALIBRATION OF THE PROBE TEMPERATURE.
Page 39
ADC. It may also occur if the thermocouple should open. This message is displayed when the input voltage to the ADC -03- has fallen below the lower negative limit of the ADC. Teledyne Analytical Instruments...
Page 40
Operation Model 9060L Teledyne Analytical Instruments...
Maintenance Maintenance Aside from normal cleaning and checking for leaks at the probe connection, the Model 9060L should not require any maintenance beyond replacement of a blown fuse. Routine maintenance includes occasional recalibration, as described in Chapter 4, Operation .
Page 42
Maintenance Model 9060L 3. Insert a small flat-blade screwdriver into the slot in the receptacle wall nearest the fuse and gently pry open the fuse receptacle. The fuse holder will slide out. There are two fuses in use and are visible in the clip.
Response Time : 90% in less than 4 seconds at 25 °C. System Power Requirements : 120 VAC, 50/60Hz (optional 220VAC), 110W. System Enclosure : Panel mount. Dimensions (Approx.) 7.0 W x 6.75 H x 4.26 D. Sensor Type : Zirconia, solid state Teledyne Analytical Instruments...
Notes: 1. Care must be taken to avoid contact with explosive or inflammable gases with 9060L heated probes when hot. 2. Please contact factory for corrosives other than sulphur or zinc. We can provide test materials to try in your atmosphere.
D-77712 Outline Diagram D77713 Control Unit Final Assembly C77925 Interconnection Diagram A.5 Probe or Sensor EMF Tables ZIRCONIA OXYGEN SENSOR OUTPUT (mV) PROBE TYPE 9060L, SENSOR TYPE 9060LEX % OXYGEN mV at 720°C (1320°F) 21.0 0.00 20.5 0.46 20.0 0.99 19.5...
Page 47
'K’ TC mV 29.212 at 720 °C (1320°F) These tables are based on the Nernst equation: Sensor e.m.f. = 0.02154 x T x 1n ( 20.95 / % oxygen), where T = ° K (° C + 273), e.m.f. is in mV’s Teledyne Analytical Instruments...
Page 48
Appendix Model 9060L Teledyne Analytical Instruments...