Teledyne 9060L Operating Instructions Manual page 47

Percent oxygen analyzer
Table of Contents

Advertisement

Percent Oxygen Analyzer
11.0
10.5
10.0
9.5
9.0
8.5
8.0
7.5
7.0
6.5
6.0
5.5
5.0
4.5
4.0
3.5
3.0
2.5
2.0
1.5
1.0
0.5
0.2
'K' TC mV 29.212 at 720 °C (1320°F)
These tables are based on the Nernst equation:
Sensor e.m.f. = 0.02154 x T x 1n ( 20.95 / % oxygen),
where T = ° K (° C + 273), e.m.f. is in mV's
Teledyne Analytical Instruments
13.78
14.78
15.82
16.92
18.08
19.30
20.60
21.98
23.45
25.04
26.75
28.61
30.65
32.90
35.42
38.28
41.58
45.48
50.25
56.41
65.08
79.91
99.51
Appendix