Ac Current Transformer (Y8101) - Fluke 8062A Instruction Manual

True-rms multimeter
Table of Contents

Advertisement

8062A
Instruction Manual
around conductors up to 3/4 inch in diameter. The pistol shape allows safe,
easy, one-hand operation when making current measurements.
The Model Y8100 probe is battery powered with size AA cells. It measures
current to 200A dc or ac rms using most any voltmeter. Two ranges, 20A and
200A, produce a 2V output at full-range current.
RANGES...........................................
RATED OUTPUT.............................
ACCURACY
DC to 200 Hz ..................................
200 Hz to 1 kHz ..............................
CALIBRATION CYCLE..................
FREQUENCY RESPONSE ..............
RECOMMENDED LOAD................
TEMPERATURE RANGE ...............
HEATING LIMITATION.................
WORKING VOLTAGE RATING....
APERTURE SIZE.............................
SIZE OVERALL...............................
WEIGHT...........................................
POWER.............................................
BATTERY LIFE...............................

6-11. AC Current Transformer (Y8101)

The Model Y8101 is a small clamp-on current transformer designed to
extend the current measuring capability of an ac current meter up to 150
amperes. A clamp-on coil designed into the probe allows measurements to
6-10
20A ac or dc
200A ac or dc
2V at full range
2% of range
<100A add 3% reading
>100A add +6% reading
1 year
dc to 1.0 kHz
3.0 k
+15 C to +35 C; for specified
accuracy -10 C to +50 C; storage
and operation at reduced accuracy.
Prolonged operation above 200A ac
or 1 kHz can cause damage to the
Y8100.
Core to output; 600V dc or 480V ac
maximum output to ground; 42V dc
or 30V ac Max.
3/4" (19 mm) diameter
9" x 4-1/2"x 1-7/16" (230 mm x 115
mm x 37 mm)
14 ounces (0.4 kg), with batteries
Battery, four AA cells
Alkaline-20 hours continuous