Fluke 8062A Instruction Manual page 20

True-rms multimeter
Table of Contents

Advertisement

8062A
Instruction Manual
Continuity
Ranges........................................... All resistance ranges
Threshold ....................................... Nominally <50% of range (for
Display Indication........................... Horizontal bar across the top of the
Response Time.............................. 50 s maximum (10 s typical)
Overload Protection ....................... 300V dc or rms ac
Diode Test
Range ............................................ 2V
Test Current ................................... 1 mA (typical)
Accuracy ........................................ (0.06% of reading + 2 digits)
Response Time.............................. 1 seconds maximum
Overload Protection ....................... 300V dc or rms ac
DC Current
Range
Resolution
200 A
2 mA
20 mA
200 mA
2000 mA
Overload Protection ....................... 2A/250V fuse (operator replaceable)
1-8
Table 1-2. 8062A Specifications (cont)
(% of reading + no. of digits)
.01 A
.1 A
1 A
10 A
100 A
example, 100
in the 200
for 200 , 2 k , 20 k , 200 k
ranges. Nominally <100 k in
autoranging k .
display and/or 2.667 kHz tone.
Indication is present for a minimum of
200 ms.
(Specification applies for voltage
measurement)
Accuracy
0.3% + 2
0.7% + 2
in series with 3A/600V fuse (service
personnel replaceable).
range)
Burden
Voltage
.3V typical
.3V typical
.3V typical
.3V typical
.9V typical